SensoCard Test
High quality blood glucose test strip
|
The SensoCard Test provides an easy, accurate method for the determination of capillary whole blood glucose content. These blood glucose test strips were designed to be used with the SensoCard and SensoCard Plus blood glucose meters.
Accuracy :
Accuracy of SensoCard Test strip was assessed by comparison of patients with diabetes in one tipical clinical evaluation. Reference measurements were done by another approved blood glucose monitor. Specimens ranged from 1.0 mmol/l to 26.4 mmol/l, as measured by the reference method. The linear regression data from one tipical clinical site study is presented in the table below.
Number of Samples: | 198 |
Range (mmol/l): | 1.2 - 27.8 |
Bias (%): | -4.45% |
Slope: | 1.055 |
Intercept (mmol/l): | -0.7182 |
Correlation Coefficient: | 0.994 |
A laboratory study (test based on Standard EN ISO 15197:2003) was conducted with SensoCard Test strip using fresh sodium EDTA venous blood with a broad range of glucose levels. Results from a single typical run of this study, shown in the table, exhibit excellent precision characteristics
for SensoCard Test strip. From all the studies taken together, estimates of system precision were calculated within-run precision, 3.65% between run precision, 3.42%.
Individual Use Method
Number of Reading: | 100 | 100 | 100 | 100 | 100 |
Average (mmol/l): | 2.9 | 4 | 6.9 | 11.4 | 17.1 |
SD (mmol/l): | 0.12 | 0.15 | 0.22 | 0.35 | 0.68 |
CV (%): | 4.1 | 3.75 | 3.2 | 3.1 | 4.0 |
Technology: |
advanced biosensor |
Sample volume: |
0.5 μl |
Test time: |
average 5 seconds |
Measurement range: |
20 – 600 mg/dl (1.1-33.3 mmol/l) |
Display alarms: |
Low: below 20 mg/dl (1.1 mmol/l) |
Hematocrit range: |
30 - 55 % |
Storage (temperature): |
8 - 30 °C |
Storage (relative humidity): |
30 - 70% |
-
10, 25 or 50 strips in a vial
-
Code-card
-
Instructions for use
![]() |
Download SensoCard product leaflet |
![]() |
Download SensoCard Test user manual |